|
Experiments with Silane |
General Safety
Precautions.
Always
wear safety glasses. Gases in syringes may be under pressure and
could spray liquid chemicals. Follow the instructions and only use
the quantities suggested.
Use a fume
hood if available.
The gas-generation
and gas-washing steps should be carried out inside a working fume hood.
Toxicity.
Silane is spontaneously flammable in air and produces SiO2
when it burns. Do not breath airborne SiO2 fumes
which are known to cause silicosis disease in the lungs. These experiments
must be done in a fume hood.
Suitability.
All of these experiments are suited for use as classroom demonstrations.
The techniques described herein are more advanced than those used in the
first ten parts of this series. Individuals attempting these experiments
should be experienced with the simpler syringe/gas techniques. These
experiments are not generally advised for use as laboratory experiments
conducted by typical high school students. Advanced students or students
with special laboratory skills could be allowed to perform these experiments
under close supervision by the instructor.
Syringe Lubrication.
We recommend
lubricating the black rubber diaphragm of the plunger with silicone spray
(available from hardware stores) or medium-grade silicone oil (Educational
Innovations, $5.95 Part #GAS-150; Fisher Catalog Number S159-500; $34/500
mL.)
Equipment. (This equipment can be ordered from a variety of vendors including Educational Innovations, Flinn Scientific (US sales only), Micro Mole, and Fisher Scientific. Part numbers and links to their websites are provided.)
All of these experiments can be done with one or two syringefuls of silane.
Preparation
of Silanes.
Silane is generated by the reaction of magnesium silicide, Mg2Si(s)4, with
1 M HCl(aq) by the In-Syringe Method.
Thoroughly lubricate the diaphragms on two plungers with silicone oil or
spray so that the plungers move readily in the syringe barrels. Place
0.20 g Mg2Si(s) in a vial cap and then lower the cap
into the syringe barrel by water flotation. The cap should rest upright
on the bottom of a syringe with all of the reagent still in the cap.
Install the plunger while maintaining the syringe in a vertical position.
The plunger should fit snugly against the rim of the vial cap. Pour
approximately 15 mL 1 M HCl(aq) into a small weighing boat. Draw
10-mL of the HCl into the syringe and then cap the LuerLOK fitting.
Shake the syringe in order to react the chemicals. Note! You
will notice a few flashes of fire inside the syringe as silane is generated
and reacts with the small amount of air originally present in the syringe.
The theoretical yield for this reaction is 64 mL, however, no more than
50-mL SiH4(g) are typically produced. After
the reaction has stopped, it is necessary to transfer this highly flammable
gas to a clean, air-free syringe.
Transferring
Silane to a Clean Syringe.
Silane is a pyrophoric gas; exposure to air will result in instant fire!
Without air present, silane will not burn. Equip a clean syringe
with a 15-cm length of latex tubing. Suction up 15-mL water and invert
the syringe so that the tubing is facing upward. Discharge 5-mL water
to remove bubbles of air. Use a hemostat (or clamp) to close the
latex tubing near the end so that the tubing is filled with water as shown
in Figure 1.
Connect the latex tubing to the silane-preparation syringe as follows. Position the two syringes with the water-filled syringe on top and the silane preparation syringe on the bottom as shown in Figure 2. Use a ring stand and two clamps (not shown) to hold the syringes in position. CAUTION! Remove the syringe cap from the silane-preparation syringe with a pliers; the contents are likely to be under positive pressure and some flashes of fire with popping noises may be noted. The pliers serves to keep hands a safe distance away. Caution: Remove the syringe cap with the cap end of the syringe directed upward so that the reagents will not spray out of the syringe.
After the two syringes are connected as per Figure 2, transfer the silane to the water-filled syringe by pushing inward on the lower plunger at the same time as pulling outward on the top plunger at the same rate. When silane transfer is complete, transfer about half of the water (5-mL) from the upper syringe into the lower syringe. This fills the latex tubing with water. Clamp the tubing shut near the lower end (nearest the silane preparation syringe) so that the tubing remains filled with water. Remove the silane preparation syringe from the latex tubing and set it aside.
Figure 1 |
Figure 2 |
"Store" silane as shown in Figure 3. Connect the latex tubing to a glass pipet (the diameter of the tubing is such that it fits snugly inside the glass pipet.) The lower elbow of the latex tubing is filled with water and serves as an "air-lock" for the silane. Use a ring stand and two clamps (not shown) to hold the syringe and the glass pipet in this position. Remove the hemostat after the apparatus has been clamped in this position.
Figure 3
Disposal.
Unwanted samples of SiH4(g)
should be discarded by burning in a fume hood as per Experiment 1A or by
reaction with KOH(aq) as per Experiment 5.
Experiment
1. Silanes React with Air.
Equipment:
Part B.
Place a clean, dry medium-size
test tube over the end of the pipet. Discharge about 10 mL of silane
into the test tube . A white coating of SiO2(s)
will be deposited on the walls of the test tube as the gas burns.
As the oxygen becomes depleted, the silane will dissipate to the mouth
of the test tube and form a white cloud without flames.
Part C.
Discharge some silane through
water in a beaker or plastic cup. It is necessary to temporarily
reposition the pipet for this procedure. Silane does not dissolve
in water and the bubbles rise to the surface where they burst into flames
upon contact with the air. Suction 5-mL water back into the latex
tubing in order to re-establish the airlock. Return the pipet to
its "stored" position as shown in Figure 3.
Experiment
2. Silane Reacts with Oxygen.
Equipment:
SiH4(g) + 2 O2(g)
SiO2(s) + 2 H2O(l)
Figure 4
Experiment
3. Silane Reacts with Chlorine.
Equipment:
SiH4(g) + 4 Cl2(g) -----> SiCl4(l) + 4 HCl(aq)
A more impressive reaction occurs if the Cl2(g) is delivered directly into the silane gas with a 25-cm length of latex tubing that has been worked up to the gas phase of the silane-containing test tube. Purge the latex tubing of air before use.
The tetrachlorosilane produced by the reaction reacts with water to produce Si(OH)4 and HCl(aq).
Reaction between chlorine and
silane.
Experiment
4. Thermal Decomposition.
Equipment:
SiH4(g) -----> Si(s) + 2 H2(g)
Experiment
5. Reaction with KOH(aq).
Equipment:
SiH4(g) + (2+n) H2O(g) -----> SiO2.n H2O(s)+ 4 H2(g)
The reaction is not spectacular as the previous reactions.
This experiment should be done last because it will destroy all of the silane remaining in the syringe. Using the silane syringe/pipet-device shown in Figure 3, suction 1-mL 6 M KOH(aq) into the silane-containing syringe for every 2-mL silane. Next, suction about 10-mL water up through the pipet in order to remove KOH from the pipet and latex tubing. Close the latex tubing with the hemostat. Vigorously shake the contents for about one minute. Discharge the liquid solution. The remaining gas can be tested for silane by discharging some into the air. It should not spontaneously burn upon contact with air as it did in Experiment 1. Discharge some of the gas into a candle flame and it will burn as is expected for hydrogen.
The gas can be further
tested for hydrogen by pulling an equal volume of air into the syringe
(no fire) and then discharging the mixture onto the surface of a small
piece of activated platinum sponge (Aldrich 26715-5). (To activate
the platinum sponge, heat the sponge to red heat with a burner flame.
Allow the platinum to cool to room temperature before using in the reaction.)
The platinum will glow red as it catalyzes the combustion of hydrogen to
make water vapor.
At
the end of the experiments, wipe excess lubricant off of rubber diaphragm.
Clean all syringe parts (including the diaphragm), caps and tubing with
soap and water. Use plenty of soap to remove oil from the rubber
seal. This extends the life of the plunger. It may be necessary
to use a 3-cm diameter brush to clean the inside of the barrel. Rinse
all parts with distilled water. Be careful with the small parts because
they can easily be lost down the drain. Important: Store plunger out of
barrel.
| This article first appeared
in Chem13 News in January, 2000. The authors of the original
Chem13 article are:
From the Department of Chemistry, Creighton University, Omaha, Nebraska 68178 USA: Bruce Mattson*, faculty member, principal investigatorAlso from Creighton University: Maneesh Bansal, undergraduate student, health administration major, plans to graduate in May, 2001 and attend medical school*Author to whom correspondence should be addressed. E-Mail: xenon@creighton.edu |
Return to Contents
Go to Mattson Home Page
(This page last updated 29 January 2002)